The DUT should alarm based on the following conditions:
• If the Resistance is too low, a CQM alarm should occur.
• If the Resistance is too high, a CQM alarm should occur.
• If the Resistance changes some percentage above or below the baseline resistance, a CQM alarm should occur.
Operation Instructions
Connect the DUT Return electrodes to the CQM-2000 Test Jacks.
Set the Resistance using the knobs on the CQM-2000 Front Panel.